AMS Laserscope Greenlight HPS - Internal & External Connections and Pinouts Directed Energy Systems Last Updated: 9/26/2026 CANBUS BAUD RATE: 500,000 CHILLER RS-485 BAUD RATE: 38400 CANBUS PATH: DISP J300 -> LPS J310 -> LPS J300 -> RPB J310 -> RPB J300 -> DLB J310 -> DLB J300 -> RESONATOR PDB: Power distribution board RPB: Rear panel board DLB: Data logging board LPS: Laser power supply DISP: UI touchscreen LCB: Laser control board / resonator control board RESONATOR: Laser head assembly and LCB (resonator control board) ########################################################################### RPB J600 - Foot switch connection (Amphenol 9-pin twist-lock) 1 GND (CHASSIS) 2 STANDBY/READY PEDAL (NO) 3 STANDBY/READY PEDAL (COM) 4 YELLOW VAPOR PEDAL (NC) 5 YELLOW VAPOR PEDAL (COM) 6 YELLOW VAPOR PEDAL (NO) 7 BLUE COAG PEDAL (NC) 8 BLUE COAG PEDAL (COM) 9 BLUE COAG PEDAL (NO) RPB J700 (DB-25) - Chiller interface 1 GREEN DIGITAL GROUND 2 BLACK RS-485 B 3 RED RS-485 A 5 WHITE/YELLOW REMOTE COM 6 BLUE REMOTE IN (~7V DC to enable) 24 BROWN/BLUE/BLACK SPARE COM 25 BROWN/RED/BLACK SPARE IN RPB J200 - 24V DC power input RPB J210 - DC power connection to PDB RPB J620 - External interlock RPB J630 - Autoselect board monitoring RPB J710 - Diode stack flow switch RPB J711 - E-STOP interlock RPB J610 - E-STOP interlock chain (w/ diode driver) RPB J720 - Fiber digital interface module RPB J801 - Chassis fan RPB J1000 - Q-switch driver power output RPB J1005 - Q-switch driver fan RPB J1010 - Q-switch driver status control RPB J1030 - Q-switch driver logic control RPB J1032 - Q-switch crystal thermal monitoring RPB J1100 - Input power current transformer RPB J1500 - RESONATOR control board power output RPB J1501 - DLB power output RPB J1502 - DISP power output RPB J150X - DC power output (RESONATOR, DLB, DISP, spare) 1 +24V DC 2 GROUND 3 +15V DC 4 N/C 5 +7V DC 6 GROUND 7 N/C 8 GROUND RPB J300 & J310 (DB-15) - CANBUS & RS-485 loop to LPS and DLB DLB J200 - Power input DLB J300 & J310 (DB-15) - CANBUS & RS-485 loop to RPB and RESONATOR 1 CAN_HIGH 2 +VCAN (5V) 3 RS-485 TRANSCEIVER #1 (U310) B 4 RS-485 TRANSCEIVER #3 (U312) A 5 RS-485 TRANSCEIVER #4 (U313) B 6 CAN_LOW 7 GND 8 RS-485 TRANSCEIVER #2 (U311) B 9 RS-485 TRANSCEIVER #3 (U312) B 10 N/C 11 GND 12 RS-485 TRANSCEIVER #1 (U310) A 13 RS-485 TRANSCEIVER #2 (U311) B 14 RS-485 TRANSCEIVER #4 (U313) A 15 N/C PDB J210 - DC power connection to RPB PDB J200 - AC power input PDB J230 - AC power output to diode driver PDB J240 - AC power output to low-voltage power supply and chiller LCB J200 - DC power connection to RPB (passes through sealed twist-lock connector) 1 +24V DC 2 GROUND 3 +15V DC 4 N/C 5 +7V DC 6 GROUND 7 N/C 8 GROUND LCB J310 (DB-15) - CANBUS & RS-485 loop to DLB LCB J510 - N/C LCB J600 - Crystal heaters and thermistors 1 CRYSTAL THERMISTOR 2 CRYSTAL THERMISTOR (GND) 3 CRYSTAL THERMISTOR 4 CRYSTAL THERMISTOR (GND) 5 CRYSTAL TEC1 - 6 CRYSTAL TEC1 + 7 CRYSTAL TEC0 - 8 CRYSTAL TEC0 + LCB J1000 - Safety shutter 1 OPTO POSITION SENSOR A OUTPUT (NO) 2 OPTO POSITION SENSOR B OUTPUT (NC) 3 +COIL (WHITE) [24V DC] 4 N/C 5 -COIL (BLACK) [GND] 6 N/C 7 OPTO POSITION SENSOR EMITTER GND 8 OPTO POSITION SENSOR EMITTER +5V DC 9 NTC THERMISTOR 10 NTC THERMISTOR LCB J1010 - Aim beam board 1 +5V DC 2 ENABLE/CONTROL (HIGH/0-5V) 3 GROUND 4 N/C LCB J1100 - Feedback photodiodes (PIN-3CDI) 1 UPPER PHOTODIODE (BLACK) [PD CATHODE] 2 UPPER PHOTODIODE (WHITE) [PD ANODE] 3 UPPER PHOTODIODE (SHIELD) [PD CASE] 4 LOWER PHOTODIODE (BLACK) [PD CATHODE] 5 LOWER PHOTODIODE (WHITE) [PD ANODE] 6 LOWER PHOTODIODE (SHIELD) [PD CASE] LCB J1400 - N/C FIBER INTERFACE CONNECTOR AND ID MODULE 0105-3730 [ALIGNMENT FIBER BRIDGES 5 TO 6 and 9 TO 10] BLACK: TEMPERATURE TRANSDUCER (AD590KF) PURPLE: TEMPERATURE TRANSDUCER (AD590KF) BROWN: CONTACT #12 ORANGE: CONTACT#1 GREEN: CONTACT #2,3,6,7,10,11 RED: CONTACT #4 YELLOW: CONTACT #5 WHITE: CONTACT #9 BLUE: CONTACT #8